Our pursuit and enterprise aim would be to "Always fulfill our buyer requirements". We carry on to acquire and layout excellent quality items for the two our old and new clients and realize a win-win prospect for our shoppers in addition as us for Hospital High Flow Oxygen Machine, Vet Oxygen Machine, 5l Oxygen Concentrator, For further queries or should you have any question regarding our products, please do not hesitate to contact us.
factory Outlets for Black Medical Auto Hemoglobin Analyzer - Microcuvette for Hemoglobin Analyzer – Merry Detail:

Microcuvette for Hemoglobin analyzer

 

Microcuvette for Hemoglobin analyzer0

 

Hemoglobin analyzer microcuvette

 

Product Details:

◆Material: polystyrene

◆Shelf Life: 2 years

◆Storage temperature: 2°C35°C

◆Relative Humidity≤85%

◆Weight: 0.5g

◆Packing: 50 pieces/bottle

Positive value/Reference range Reference Range:

◆Adult males: 130-175g/dL

◆Adult females: 115-150g/dL

◆Infant: 110-120g/dL

◆Child: 120-140g/dL

Test Result

◆The measurement display range is 0-250g/L. Coagulation may cause the blood specimen to fail to fill the microcuvette, resulting in incorrect measurements.

◆Hemolysis can affect the test results。

Limitation of Test Method

◆Diagnosis and treatment should not rely solely on the test result. Clinical history and other laboratory tests should be considered

Performance Specification

◆Blank1g/L

◆Repeatabilitywithin range 30g/L to 100g/L, SD3g/L; within range 101g/L to 250g/L, CV1.5%

◆Linearitywithin range 30g/L to 250g/L, r0.99

◆AccuracyThe correlation coefficient (r) of the comparison experiment is 0.99, and the relative deviation is 5%

◆Inter-batch difference≤5g/L

Test Procedure EDTA blood Testing:

◆Stored samples should be returned to room temperature and mixed thoroughly before testing.

◆Use a micropipette or pipette to draw no less than 10μL of blood on a clean glass slide or other clean hydrophobic surface.

◆Using the tip of the reagent to contact the sample, the sample enters under capillary action and fills the reagent piece.

◆Carefully wipe off any excess sample on the surface of the microcuvette.

◆Place the microcuvette on the microcuvette holder of hemoglobin analyzer and then push the holder into the analyzer to start the measurement.


Product detail pictures:

factory Outlets for Black Medical Auto Hemoglobin Analyzer - Microcuvette for Hemoglobin Analyzer – Merry detail pictures

factory Outlets for Black Medical Auto Hemoglobin Analyzer - Microcuvette for Hemoglobin Analyzer – Merry detail pictures

factory Outlets for Black Medical Auto Hemoglobin Analyzer - Microcuvette for Hemoglobin Analyzer – Merry detail pictures


Related Product Guide:

We've been committed to offering easy,time-saving and money-saving one-stop purchasing service of consumer for factory Outlets for Black Medical Auto Hemoglobin Analyzer - Microcuvette for Hemoglobin Analyzer – Merry, The product will supply to all over the world, such as: Nepal, Libya, Brazil, we rely on own advantages to build a mutual-benefit commerce mechanism with our cooperative partners. As a result, we've gained a global sales network reaching the Middle East, Turkey, Malaysia and Vietnamese.
  • This manufacturer can keep improving and perfecting products and service, it is in line with the rules of market competition, a competitive company.
    5 Stars By Abigail from Angola - 2018.11.22 12:28
    Good quality and fast delivery, it's very nice. Some products have a little bit problem, but the supplier replaced timely, overall, we are satisfied.
    5 Stars By Rosalind from Florida - 2018.03.03 13:09

    Related Products