2IBP

Short Description:

◆ 2 channels of invasive blood pressure.

◆ Simultaneous measurement of Systolic, Diastolic and Mean pressure.

◆ The product is a high accuracy blood pressure measure equipment which is used with corresponding monitor. It can measure the blood pressure of the chosen vas (SYS / MAP / DIA). It is fit for the blood pressure monitor of adult, children and infant.


Product Detail

2IBP for patient monitor

 

2IBP
2IBP

Purpose:

◆With Template self-adapting technique, which ensure the accuracy in complicated clinic conditions. Solve the problem that NIBP of most brand patient monitor cannot measure stably and accurate when patients have arrhythmia and rapid changing pulse. The anocelia and intracranial will be monitored at the same time. It is very suitable to the monitor of critically ill patients.

 

Physics Discreteness

IBP cable, IBP sensor, IBP module

Features

◆Plug and play IBP technique

◆It is fit for adult, enfant and infant.

◆ Dual-channel measure

◆Low-power consumption, dispense with special power supply

◆Automatic digital data processing

◆With time after time zero calibration, calibration function

2IBP module can monitor parameters as follows

◆ART          Arterial Blood Pressure
◆PA             Pulmonary Arterial Pressure
◆CVP           Center Venous Pressure
◆RAP           Right Atrial Pressure
◆LAP           Left Atrial Pressure
◆ICP            Intracranium Pressure
◆P1-P2         Expand Pressure
Monitor function

◆CH1:SYS(Systolic pressure) DIA(Diastolic pressure) MAP(Mean pressure)
◆CH2:SYS(Systolic pressure) DIA(Diastolic pressure) MAP(Mean pressure)

Parameter

Measure range : ART                                 0~40(KPa)
PA                                  -0.8~16(KPa)
CVP/RAP/LAP/ICP          -1.3~5.3(KPa)
P1—P2                           -1.3~40(KPa)
◆Zero Calibration :every time

◆Calibration: Termly calibration or the time when using the new sensor
◆Sensitivity:5μV/V/mmHg
◆Impedance range:300~3000(Ohm)
◆Resolution:1(mmHg)
◆Accuracy:±2% or 1mmHg (select larger)
◆Update Interval: about 1 Sec
◆Display mode: corresponding monitor screen

◆Label: ART, AoP, LVP, PAP, CVP, WDG, RAP, RVP, ICP, IAP and IEP.
◆unit: mmHg.
◆CVP, ICP, IAP and IEP can measure in mmHg or cmH20.

re